High Vacuum Block Valve - Angle Valve Series
Jul. 22, 2020
High Vacuum Block Valve - Angle Valve Series
This series of valves is divided into manual, pneumatic and electromagnetic driven types. Features with smooth operation, small size, reliable use, good sealing performance and long service life. It is one of the preferred valves for vacuum equipment. The valve respectively rotates the handle, the compressed gas pushes the cylinder and the coil energizes to generate electromagnetic force, and the power is connected to the valve plate through the mechanism, and drives the valve plate to open and close. Applicable medium can be pure air and non-corrosive gas.
Technical Parameters | ||
Application scope | 1x10^(-5)Pa~1.2x10^(5)Pa(Viton o-ring sealed) 1x10^(-6)Pa~1.2x10^(5)Pa(Bellow seal) | |
Differential pressure on the gate | DN 200~250(Pneumatic Type):≤3x10^(4)pa
The other types:≤1.0x10^(5)Pa Any direction | |
Body and seat leak rate |
1.3x10^(-7)Pa.L.S^(-1) | |
Cycles until first service |
200,000 times | |
Bake-out temperature | ≤150℃ | |
Installation position | Manual drive: any direction Pneumatic drive: any direction (DN ≤ 160); Sealed surface direct to the vacuum Electromagnetic drive:any direction (bellow sealing); Sealed surface direct to the vacuum(Viton O-ring sealed) | |
Compressed air (only supply to pneumatic valves) |
0.4-0.7MPa | |
Power Supply | Pneumatic valves: AC 220V 50HZ, 6W or DC 24V, 3W( can be customized) Electromagnetic drive: AC 220V 50HZ | |
Speed of opening or closing: | Pneumatic drive:DN16~DN50≤1s;DN63~DN250≤3s Electromagnetic drive: open≤0.1s;close≤1s(viton sealing) close≤0.5s(bellow sealing) | |
Valve position indication: | Manual drive: with open&close position indication Pneumatic drive: with open&close position indicator switch(magnetic switch) Electric drive: with turn-on indicator |
Optional Accessories
Pneumatic DN 16~50: open and close position indicating switch (magnetic switch), Model D-A73; Pneumatic DN 63~250:ISO flange; two-position five-way electromagnetic reversing valve, Model 4V210-06.
Precautions in use:
1. The valve should first check whether the valve is intact and the accessories are complete.
2. The valve should be kept clean and stored in a dry room and protected from strong vibrations.
3. When the valve is not used for long-term storage, the valve should be in the micro-open state, and should be checked regularly to prevent moisture, rust and rubber parts aging.
4. Before installation, the surface of the valve and vacuum should be cleaned according to vacuum hygiene requirements.
5. The user's flange connected to the valve must have no protruding welds in the joint hole.
6. The air supply pressure to the pneumatic valve should be guaranteed in the range of 0.4 to 0.7 MPa.
7. The signal switch of the pneumatic valve is a passive switch signal, and the maximum load of the switch is 0.1A 250VAC.
8. The newly used pneumatic valve may encounter a higher pressure of the air source required for shutting down. This is a normal phenomenon, and the pressure of the air source required to close after the number of times of operation is decreased.
9. The interval between two operations of the electromagnetic valve should not be less than 1 s, and the power supply should be reliable, and there should be no instantaneous on-off phenomenon, otherwise the valve could not be opened.
10. In the solenoid valve power supply circuit, the fuse is connected in series. For short-circuit protection, the recommended fuse current is 1A (DN16), 2A (25), 3A (DN 40 and 50).
The possibility of fault and eliminating methods
Fault | Reason | Eliminating methods |
Weak sealed | Seal face with oil | Clean the oil spot |
Seal face has been nicked | polish the nick with polishing paper or the machine | |
Rubber sealing ring has been damaged | Change rubber sealing ring | |
Bellow has been damaged | Change or weld the bellow | |
Can not open/close (pneumatic) | Insufficient supply pressure | Raise the supply pressure to standard level |
Throttle valve with a wrong position | Reset to initial state | |
Air pressure unbalanced | Adjust to be balanced | |
Reversing valve fault (Pneumatic) | Insufficient supply pressure | Raise the supply pressure |
Reversing valve without power | Power on | |
Reversing valve has been damaged | Change the reversing valve | |
Can not open/close (electromagnetic) | Out of the range of power supply | Adjust to the certain range |
poor contact for loose or break connection | Change the broken wire or connect again | |
Short-circuit damage of circuit board | Return to the factory for renewal |
-